On Call Sure clinical accuracy

EN ISO 15197:2015 acceptance criteria for system accuracy:

95% values shall fall within either ± 0.83 mmol/l (± 15 mg/dl) of the average measured values of the reference measurement at glucose concentrations < 5.55 mmol/l (< 100 mg/dl) or

95% within ± 15% at glucose concentrations ≥ 5.55 mmol/l (≥ 100 mg/dl)

On Call Extra Mobile results:

100% within either ± 0.83 mmol/l at glucose concentrations < 5.55 mmol/l

98.8% within ± 15% at glucose concentrations ≥ 5.55 mmol/l

On Call Sure results:

100% within either ± 0.83 mmol/l at glucose concentrations
< 5.55 mmol/l

100% within ± 15% at glucose concentrations ≥ 5.55 mmol/l

Confidence in pregnancy

Haematocrit can change rapidly during pregnancy. Because accuracy is not significantly affected by varying haematocrit when testing blood glucose levels using the On Call  meters. The On Call Extra Mobile and On Call Sure Sync are perfect blood glucose meters for patients with gestational diabetes.

On-Call-Extra-Mobile

Product specifications

For full product specifications on the On Call Extra Mobile view the spec sheet below:

View the On Call Extra Mobile spec sheet

Product Evaluation Paper

For accuracy information about the On Call Extra Glucose Meter, view the evaluation paper below:

Evaluation of On Call Extra Analytical Performance to EN ISO 15197:2015 Standard

On Call Sure

Product specifications

For full product specifications on the On Call Sure Sync view the spec sheet      below:

On Call Sure Sync view the spec sheet

Product Evaluation Paper

For accuracy information about the On Call Sure Glucose Meter, view the evaluation paper below:

Evaluation of On Call Sure Analytical Performance to EN ISO 15197:2015 Standard